Cycle Sterilization Temp. Sterilization Time Total Cycle Time
(*does not include drying)
STATIM 2000 G4
Unwrapped 135°C 3.5 Minutes 6 Minutes
Wrapped 135°C 10 Minutes 14 Minutes
Rubber/Plastic 121°C 15 Minutes 20 Minutes
Heavy Duty Unwrapped
/Air Drying Air drying only Air drying only Air drying only
STATIM 5000 G4
Unwrapped 132°C 3.5 Minutes 9 Minutes
Wrapped 132°C 6 Minutes 17.5 Minutes
Rubber/Plastic 121°C 35 Minutes 45 Minutes
Heavy Duty Unwrapped
/Air Drying 132°C 6 Minutes 12 Minutes
STATIM 2000 G4
Unit Size (L x W x H) 19.5” x 16.3” x 5.9” or 49.5 cm x 41.5cm x 15cm
Cassette Internal Dimensions 11” x 7.1” x 1.4” or 28cm x 18cm x 3.5cm
Weight Without Water 46lbs or 21kg
Power Consumption 110V, 60Hz, 11A
Benefits of the G4 network:
Monitoring real-time cycle data
Instant data logging
Proactive maintenance management
Automatic system updates to your G4 machines
FAST.
STATIM‘s unmatched speed and dependability have made STATIM the
world’s best-selling autoclave. And its patented gentle
technology has made STATIM the world’s favorite.
FORWARD.
The STATIM G4 will offer you a level of interactivity never
seen before. The STATIM G4, through its LCD touch screen control
and STATIM Live networking capability, has been designed to provide
you with unprecedented ease of use and puts the user in total control.
DOCTORS AROUND THE WORLD RELY ON STATIM.
Healthcare practitioners in over 90 countries around the world rely
on STATIM to protect their patients and staff from infection –
with sterilized instruments in just minutes. In 1989, SciCan transformed
the industry, developing the world’s first compact sterilization unit
designed to meet the growing need for handpiece sterilization. Today,
the STATIM still boasts superior cycle times and processes for both wrapped
and unwrapped cycles. In fact, the STATIM Cassette Autoclave® sterilizes up
to ten times faster than traditional ‘chamber style’ units. Employing a proprietary
positive pressure pulse displacement technology – using saturated steam to remove air
and effectively sterilize instruments between patients – the STATIM Autoclave ensures
dentists always have an ample supply of instruments ready for use.
STATIM‘S SPEED
Reduces Instrument Investment
STATIM sterilizes handpieces and instruments for immediate use in as little as 8 minutes,
reducing the number required as instruments can be processed ‘just in time’ or between patients.
Increases Efficiency
Fast point-of-use sterilization makes turnaround a priority, increasing efficiency
and virtually eliminates instrument bottlenecks.
Gentle on Instruments
Our state-of-the-art steam technology is gentle on both solid and hollow instruments.
THE STATIM DIFFERENCE.
Immediate Steam Generation
Unlike conventional autoclaves that need to heat an entire reservoir of water,
STATIM uses only a small, specific amount of water required for each cycle.
Water is introduced into a heating device referred to as a ‘steam generator’,
rapidly heated and converted to steam in mere seconds.
Air Removal
Using PPPD technology and our patented steam injection process – air is
automatically forced out of the sterilization chamber.
Cassette System
The thin stainless steel walls of STATIM’s fully-removable cassette,
(or sterilization chamber) allow for rapid heating and cooling of the chamber.
As a result, STATIM meets the conditions for sterilization and drying
in a fraction of the time required by a conventional autoclave.
Drying Technology
The STATIM 2000 and 5000 have each been designed to provide you with
a complete solution: rapid sterilization balanced by rapid drying.
TOUCH. CONNECT. PROTECT. COLLECT.
Touch
A large 3.5” high-resolution touchscreen offers a vivid display of messages,
current cycle information and customizable colours all with extraordinary clarity.
Fully adjustable contrast levels let you optimize the screen
to the lighting in your steri-centre.
Connect
Through your networked portal, you can view not only your STATIMs current
operations in real time, you can also access cycle history,
printing and validation information, and also sterilization records.
Capable of sending emails with cycle information errors directly to your dealer,
your STATIM G4 will let qualified technicians know when service is imminent.
Protect
Patient safety relies on attention to detail. The STATIM G4 supports verifiable
processes that are PIN protected with user-defined IDs so that staff members can
be required to authorize a cycle. This encourages extra attention to process
and creates a safer environment for both staff and patients.
Collect
An undocumented sterilization cycle might as well not have been performed.
With each cycle, information must be stored to protect your office and
your patient. The integrated DataLogger stores all cycle information for
the lifetime of the unit, which is accessible through multiple channels such as:
a USB port for manual retrieval; an Ethernet connection for web based archiving
or desktop printing; or a RS232 port for an external printer option.
Reviews
There are no reviews yet.